Skip to content

Public FDA record

Z-0218-2019: NucliSENS Magnetic Extraction Reagents

Recall record Z-0218-2019 for BioMerieux SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0218-2019
Company in FDA record
BioMerieux SA
FDA record creation date
Oct 23, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Data loggers showed that one shipment to the United States experienced low temperatures that may impact performance of this product.

Source facts

Action
bioMerieux notified customers on about 09/17/2018 via "URGENT PRODUCT REMOVAL NOTICE". Instructions included the following. 1. Confirm this information has been distributed to, and reviewed by, all appropriate personnel within your organization. Retain a copy of this letter for your files, and forward this information to all parties that may use this product, including others to whom you may have transferred this product. 2. Please discontinue use and destroy any impacted product remaining in your local inventory and contact your local bioMrieux, Inc. representative for credit. 3. Among tests previously performed with these batches, we advise you to discuss patients results: identify any possible false results, retest impacted samples when possible, analyze the related risks and determine appropriate actions if relevant. 4. Complete the attached Acknowledgement Form in Attachment A to confirm receipt of this notice and return it to your local bioMrieux, Inc. representative. Questions and concerns can be directed to Clinical Customer Service representative at (800) 682-2666 or Industry Customer Service representative at (800) 634-7656.
Initiated
2018-09-17
Root cause
Other
Terminated
2020-05-18
Distribution
Nationwide distribution to MD, TX, MA, GA, OH, OR, CA, and WA.
Recall number
Z-0218-2019
Recall status
Terminated
CDRH record ID
167420
FDA record created
2018-10-23
Recall event number
81067

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗