Z-0218-2024: CardioMEMS Patient Electronic System (PES), Model Number CM1100. Used for wirelessly measuring and mon

Recall record Z-0218-2024 for St. Jude Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0218-2024
Company in FDA record
St. Jude Medical
FDA classification date
Nov 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.

Source facts

Action
Abbott notified clinics on about 10/04/2023 via hand delivery of IMPORTANT MEDCIAL DEVICE CORRECTION letter. Clinics were instructed to continue following the IFU and also the supplemental guidance in the letter to continue taking readings as your Health Care Professional has recommended, inspect the Power Connector Plug for damage prior to taking your next reading, if you suspect or identify damage to the power connector plug or any other issues, call Abbott Technical Support, and after the initial inspection, Abbott requests that each Patient Electronics System power connector plug is inspected monthly for damage. Clinics were also notified that patients would be notified of the issue using the addresses listed in the Merlin.net" Patient Care Network; please ensure all patient addresses are up to date. Patients were notified on about 10/18/2023, via IMPORTANT MEDCIAL DEVICE CORRECTION letter. Patients were instructed to continue to take readings as your Health Care Professional has recommended, inspect the power connecter plug for damage prior to taking your next reading, if you suspect or identify damage to the power connector plug contact Abbott Technical Support at 1-844-MYCMEMS (1-844-692-6367) Monday-Friday: 8am-8pm ET, and Abbott requests that each Patient Electronics System power connector plug is inspected monthly for damage.
Initiated
2023-10-04
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IRELAND, ITALY, KUWAIT, NETHERLANDS, PORTUGAL, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UNITED ARAB EMIRATES, UNITED KINGDOM.
Recall number
Z-0218-2024
Recall status
Open, Classified
CDRH record ID
203531
FDA classified
2023-11-03
Recall event number
93141

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗