The 10 mm/135 degree Titanium Cannulated Trochanteric Fixation Nail 170 MM (Sterile) was assembled incorrectly. If the non conformance goes undetected, it may impact the locking function post-operatively and compromise the bone reduction and construct stability which may lead to non-union/malunion.
Source facts
Action
An urgent medical device recall notice, dated October 29, 2014, was sent to end users that identified the product, problem and action to be taken. Customers were instructed to return the device to Synthes along with completing the verification section of the notice.
Initiated
2014-10-29
Root cause
Employee error
Terminated
2015-12-18
Distribution
US Distribution in the states of: AK, FL, OH, OK, and TX.
Recall number
Z-0219-2015
Recall status
Terminated
CDRH record ID
131346
FDA classified
2014-11-14
Recall event number
69685