Open original FDA record ↗Recall record
- FDA identifier
- Z-0220-06
- Company in FDA record
- Smiths Medical PM, Inc.
- FDA classification date
- Nov 30, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.
Source facts
- Action
- Distirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004. Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB). The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.
- Initiated
- 2005-09-01
- Root cause
- Other
- Terminated
- 2007-06-27
- Distribution
- AZ, CA, KS, KY, NY, TN, TX OUS to countries such as: Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain
- Recall number
- Z-0220-06
- Recall status
- Terminated
- CDRH record ID
- 41764
- FDA classified
- 2005-11-30
- Recall event number
- 33375