Open original FDA record ↗Recall record
- FDA identifier
- Z-0220-2020
- Company in FDA record
- LivaNova USA Inc.
- FDA record creation date
- Oct 25, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm has become aware that due to a chemical reaction, the disinfectant/water preservative used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. In addition, the firm is releasing a design upgrade to reduce the risk of potential emission of aerosols from the 3T.
Source facts
- Action
- On 10/18/2018, the firm sent a certified "MEDICAL DEVICE CORRECTION" Letter to its customers informing them that due to a chemical reaction, the disinfectant/water preservative (hydrogen peroxide) used to disinfect the device to inhibit/limit the growth micro-organisms, may be rendered ineffective. The letter advises customers to: -Daily Hydrogen Peroxide Monitoring: Users should monitor the hydrogen peroxide concentration in the water solution on a daily basis to verify that sufficient concentration of hydrogen peroxide is present in the water circuit of the device. A decrease in hydrogen peroxide over the 7-day-period until the next water change is expected, however the hydrogen peroxide concentration should remain above 100 ppm. Detailed instructions are provided in Attachment 1, titled Daily Hydrogen Peroxide Monitoring Instructions (included in the Customer Notification Letter). Customers are instructed to read the detailed instructions carefully and to follow them. -3T Design Upgrade: The Recalling Firm has developed a vacuum canister and internal sealing design change that is intended to further mitigate (but does not eliminate) the risk of airborne transmission of non-tuberculosis mycobacterium (NTM) from the 3T device. Important instructions that ensure the effectiveness of this modified design will be provided at the time of the upgrade. A Firm representative or local agent will contact its customers to plan the upgrade of their affected products. No other actions by the user are required at this time, other than to continue to follow the 3T Operating Instructions and the attached Daily Hydrogen Peroxide Monitoring Instructions, and to return the Customer Response Form. For those users that have devices undergoing deep cleaning, your device will be upgraded at the deep cleaning facility. Customers are instructed to please complete and return the attached Customer Response Form by fax to (303) 467-6502 or by email to USFSN@livanova.com. Cu
- Initiated
- 2018-11-01
- Root cause
- Device Design
- Terminated
- 2022-08-04
- Distribution
- Worldwide distribution - US Nationwide distribution including sate of PR and countries of Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Iran (Islamic Republic of), Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea (the Republic of), Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia (the former Yugoslav Republic of), Malaysia, Maldives, Malta, Mauritius, Mexico, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines (the), Poland, Portugal, Qatar, Romania, Russian Federation (the), Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan (Province of China), Thailand, Trinidad and Tobago, Tunisia, Turkmenistan, Turkey, Ukraine, United Arab Emirates (the), United Kingdom, Uruguay, Uzbekistan, Vietnam.
- Recall number
- Z-0220-2020
- Recall status
- Terminated
- CDRH record ID
- 176430
- FDA record created
- 2019-10-25
- Recall event number
- 83822