Z-0220-2023: VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use on

Recall record Z-0220-2023 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0220-2023
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Nov 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

Source facts

Action
Ortho Clinical Diagnostics issued Urgent Letter (CL2022-260) to VITROS 250/350 Chemistry System consignees. Letter states reason for recall, health risk and action to take: Upon receiving this communication, load Calibration Diskette DRV 6201 on your VITROS 250/350 Chemistry System and calibrate VITROS PHYT Slides, GEN 22 using VITROS Calibrator Kit 9. Complete the enclosed Confirmation of Receipt form no later than October 21, 2022. Please forward this notification if the affected product was distributed outside of your facility. Save this notification with your user documentation or post this notification by each VITROS 250/350 Chemistry System until Calibration Diskette DRV 6201 has been received and loaded by your facility. If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Ortho Care" Technical Solutions Center. If you have further questions, please contact the Ortho Care Technical Solutions Center at 1-800-421-3311
Initiated
2022-10-13
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden United Kingdom
Recall number
Z-0220-2023
Recall status
Open, Classified
CDRH record ID
196591
FDA classified
2022-11-10
Recall event number
91051

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗