Open original FDA record ↗Recall record
- FDA identifier
- Z-0221-2007
- Company in FDA record
- Computerized Medical Systems Inc
- FDA classification date
- Nov 29, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The In/Out adjustment knob can be over-turned, ultimately causing the incorrect placement of radioactive seeds in the patient.
Source facts
- Action
- The recalling firm issued letters dated 11/9/06 via regular mail. The letter informs the customer of the problem and includes an advisory notice instructing them to perform a test prior to each use. If the system does not function correctly, it can still be used safely in a manual mode until it can be returned for repair. A questionnaire was enclosed that will facilitate the firm in prioritizing repairs and retrofits. After the questionnaire is reviewed, the firm will contact the customer for further information regarding their facilities use of the device. This letter was also issued to the CMS customer support departments in Japan, China, and Europe, who will notify the foreign customers and prioritize their returns for repair.
- Initiated
- 2006-11-09
- Root cause
- Other
- Terminated
- 2007-11-30
- Distribution
- Worldwide, including USA, Australia, Japan, China, France, Germany and Brazil.
- Recall number
- Z-0221-2007
- Recall status
- Terminated
- CDRH record ID
- 49115
- FDA classified
- 2006-11-29
- Recall event number
- 36665