Z-0221-2020: CentriMag Acute Circulatory Support System Motor, U.S. Model 102956 and OUS Model 201-10002, a component

Recall record Z-0221-2020 for Thoratec Switzerland GMBH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0221-2020
Company in FDA record
Thoratec Switzerland GMBH
FDA classification date
Oct 30, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).

Source facts

Action
A recall letter dated August 22, 2019 was issued on Abbott letterhead and was delivered by field representative visits beginning August 22, 2019 for the U.S. consignees. Foreign consignees were notified with a letter dated August 22, 2019 containing the same content but OUS model numbers and OUS contact information. Depending on the requirements by the local regulator, the letters may have been delivered by a field representative or via certified mail. The letter identified the affected product, problem and actions to be taken.
Initiated
2019-08-22
Root cause
Process change control
Terminated
2024-04-26
Distribution
Distribution was nationwide. There was government distribution, but no military distribution. Foreign distribution was made to Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, France, Hong Kong, India, Israel, Italy, Liechtenstein, Mexico, Netherlands, Portugal, Qatar, Romania, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, and United Kingdom.
Recall number
Z-0221-2020
Recall status
Terminated
CDRH record ID
176695
FDA classified
2019-10-30
Recall event number
83895

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗