Skip to content

Public FDA record

Z-0221-2023: Philips Laser System, REF: LAS-100. Used in minimally invasive interventional procedures within the card

Recall record Z-0221-2023 for Spectranetics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0221-2023
Company in FDA record
Spectranetics Corporation
FDA classification date
Nov 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Laser System may detect Error 106 or Error 108 System Failures that will cause the system to enter non-recoverable safe state. The user will not be able to proceed until the error is cleared. May result in initiation of treatment delay and/or treatment beyond initial scope and/or unable to treat patient.

Source facts

Action
On 10/14/22, correction notices were mailed to customers who were asked to take the following actions: Firm recommends continued use of affected devices and to follow the operator's manual. As a temporary mitigation, until software becomes available tentatively in Q2 2023, the user can perform steps when visually alerted on the system display as follows: Error 106 Firm's internal testing confirms that the following steps may temporarily resolve the issue, if encountered: 1. System restart 2. Recalibration of the catheter in use or calibration of a new catheter Error 108 Firm's internal testing confirms that the following steps will resolve the issue, if encountered: 1. System restart 2. Recalibration of the catheter in use or calibration of a new catheter 3. Once the system is in the Ready State, pause 2 seconds before pressing the footswitch Firm recommends all affected device users within their facility are notified of correction and a copy of the correction notice should be retained. Complete and return the Customer Reply Form. Tentatively in Q2 2023, firm will contact all affected customers and arrange for a Field Service Engineer to update the system software during preventive maintenance or service visits. Distributors are asked to send the notice and response form to customers to whom affected devices were distributed and to follow-up with these customers to obtain signed response forms. Customers with additional questions are encouraged to call 1-800-231-0978 (option 1), Monday - Friday 9:00AM - 5:00PM MST, or email laserserviceops@phiIips.com
Initiated
2022-10-14
Root cause
Device Design
Distribution
US: LA, MO, TN, CA, NC, PA, HI, AZ, CO, OK, FL, TX, MN, MI, VA, CT, NY, MD, IL, NJ, MA, KS, ME, AR, IA, IN, GA, ID, OR. OUS: DE, GB, ES, SE, IT, RS, GR, IN, FR, IL, DK, SA
Recall number
Z-0221-2023
Recall status
Open, Classified
CDRH record ID
196360
FDA classified
2022-11-11
Recall event number
90971

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗