Skip to content

Public FDA record

Z-0221-2024: Arial Water Resistant Pendant with Wristband, Models: 59361 and 2560-59361; mobile patient alarm

Recall record Z-0221-2024 for Securitas Healthcare LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0221-2024
Company in FDA record
Securitas Healthcare LLC
FDA classification date
Nov 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Premature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.

Source facts

Action
Securitas Healthcare issued a recall notice to its consignees on 09/25/2023 via USPS, registered mail. The notice explained the issue, potential hazard, and requested the following actions: 1. Immediately replace Pendant battery upon Low Battery Alert 2. Customer should check Pendant battery status in Arial System application at least daily 3. Customer should check for Pendants that are no longer communicating with the Arial System at least daily This notice must be passed on to those who need to be aware within your organization. - Securitas Healthcare will be replacing all Pendants from the Affected Lots with refurbished Pendants at no cost. The notice has also been posted: https://www.securitashealthcare.com/product-advisories
Initiated
2023-09-25
Root cause
Component change control
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-0221-2024
Recall status
Open, Classified
CDRH record ID
203623
FDA classified
2023-11-03
Recall event number
93188

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗