Open original FDA record ↗Recall record
- FDA identifier
- Z-0221-2024
- Company in FDA record
- Securitas Healthcare LLC
- FDA classification date
- Nov 3, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Premature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
Source facts
- Action
- Securitas Healthcare issued a recall notice to its consignees on 09/25/2023 via USPS, registered mail. The notice explained the issue, potential hazard, and requested the following actions: 1. Immediately replace Pendant battery upon Low Battery Alert 2. Customer should check Pendant battery status in Arial System application at least daily 3. Customer should check for Pendants that are no longer communicating with the Arial System at least daily This notice must be passed on to those who need to be aware within your organization. - Securitas Healthcare will be replacing all Pendants from the Affected Lots with refurbished Pendants at no cost. The notice has also been posted: https://www.securitashealthcare.com/product-advisories
- Initiated
- 2023-09-25
- Root cause
- Component change control
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Recall number
- Z-0221-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203623
- FDA classified
- 2023-11-03
- Recall event number
- 93188