Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Source facts
Action
AD-TECH issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 09/19/2024. The notice explained the problem with the devices, risk to health, and requested the consignee contact customer support to obtain an RMA to expedite the return of the affected product.
For questions: contact your Ad-Tech Clinical Specialist.
Initiated
2024-09-19
Root cause
Employee error
Distribution
US Nationwide distribution in the states of CA, OH.
Recall number
Z-0221-2025
Recall status
Completed
CDRH record ID
210157
FDA classified
2024-10-24
Recall event number
95414
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.