Skip to content

Public FDA record

Z-0222-2020: C3 Wave App, v. 2.0.5

Recall record Z-0222-2020 for Medical Components, Inc dba MedComp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0222-2020
Company in FDA record
Medical Components, Inc dba MedComp
FDA record creation date
Oct 25, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the iPad is updated with Apple iOS software version 12, the C3 application malfunctions. Upon launching a new procedure, the ECG waves do not appear on the screen, in surface or internal mode. If the C3 application is not operating properly, the PICC procedure must be completed without using alternate methods to confirm PICC tip placement.

Source facts

Action
On October 1, 2018, the firm notified its customers via Product Advisory Notification letter. The letter communicated to customers that the most recent iOS version (12) was incompatible with the C3 Wave app v. 2.0.5. The firm released v.2.0.6 of the app on 10/3/2018, which was made available in the app store. Customers were asked to update both the iOS and the C3 app at the same time to avoid any issues within the app. Customers were provided with step-by-step directions on how to do this. If you have any questions, please contact Recalls@medcompnet.com.
Initiated
2018-10-01
Root cause
Software Manufacturing/Software Deployment
Terminated
2020-10-01
Distribution
Distributed to accounts in AZ, CA, CO, FL, IL, LA, MA, ME, NC, OH, TN, VA, and WV. Foreign distribution to Canada, Croatia, Czech Republic, Great Britain, Greece, Italy, Netherlands, Slovenia, and Sweden.
Recall number
Z-0222-2020
Recall status
Terminated
CDRH record ID
168705
FDA record created
2019-10-25
Recall event number
81439

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗