Z-0222-2025: CADD-Solis Li-ion Rechargeable Battery Packs, REF 21-2160-XX, which are also sold as a component of the f

Recall record Z-0222-2025 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0222-2025
Company in FDA record
Smiths Medical ASD Inc.
FDA classification date
Oct 31, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Damage to the battery pack may have caused a short to a capacitor within the battery pack. While the battery encasement is designed to be flame retardant, a short to the capacitor could lead to melting of the battery pack case. If this issue occurs, the battery pack charging circuit may become inoperable.

Source facts

Action
Smiths Medical issued an URGENT MEIDCAL DEVICE CORRECTION to its consignees on 07/16/2024 via traceable mailing. The notice explained the issue, potential risk and requested the following actions be taken: 1. Examine the external condition of the battery pack and look for evidence of damage to the outer case. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 2. Users with damaged battery packs should submit complaints to: globalcomplaints@icumed.com; 1-866-216-8806 3. Ensure all users or potential users of these products are immediately made aware of this notification. 4. Complete and return response form to smithsmedical5437@sedgwick.com. 5. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to smithsmedical5437@sedgwick.com
Initiated
2024-07-16
Root cause
Component design/selection
Distribution
Worldwide Distribution
Recall number
Z-0222-2025
Recall status
Open, Classified
CDRH record ID
209903
FDA classified
2024-10-31
Recall event number
95287

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗