Contact failure in F1 and F4 fuse holders was detected on the EC-5000 CXIII Excimer Laser Corneal Surgery System that could cause a serious adverse event for the patient.
Source facts
Action
Nidek Engineers visited each site to perform correction - fuse holders have been inspected and replaced if necessary.
An Engineering Change Order was issued on June 1, 2012, for inspection and/or replacement of affected fuse holders by Nidek Service Engineers.
For further questions please call (510) 353-7785.
Initiated
2012-08-06
Root cause
Component design/selection
Terminated
2014-11-17
Distribution
US Distribution including the states of CO, NY, MO, LA, ME, CA, MI, AZ, CT and FL.
Recall number
Z-0223-2015
Recall status
Terminated
CDRH record ID
130780
FDA classified
2014-11-17
Recall event number
69508