Skip to content

Public FDA record

Z-0223-2026: Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System

Recall record Z-0223-2026 for UIH Technologies LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0223-2026
Company in FDA record
UIH Technologies LLC
FDA classification date
Oct 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Source facts

Action
On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program. 5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams.
Initiated
2025-09-12
Root cause
Other
Distribution
US: OUS:
Recall number
Z-0223-2026
Recall status
Open, Classified
CDRH record ID
216305
FDA classified
2025-10-21
Recall event number
97668

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗