Skip to content

Public FDA record

Z-0224-2015: Nidek MC-500 Multicolor Laser Photocoagulator using software version 2.20. Opthalmic Device. The Nidek M

Recall record Z-0224-2015 for Nidek Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0224-2015
Company in FDA record
Nidek Inc
FDA classification date
Nov 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program.

Source facts

Action
Nidek Field Service engineers visited each affected location to upgrade software.
Initiated
2012-10-24
Root cause
Device Design
Terminated
2014-11-17
Distribution
US Distribution in the states of: NJ, SC, IL, MI, NY, CA, TX, OH, IN, UT, and MN.
Recall number
Z-0224-2015
Recall status
Terminated
CDRH record ID
130886
FDA classified
2014-11-17
Recall event number
69537

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗