The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock.
Source facts
Action
All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits.
Initiated
2023-03-13
Root cause
Component design/selection
Distribution
US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia.
Recall number
Z-0224-2025
Recall status
Completed
CDRH record ID
210236
FDA classified
2024-10-24
Recall event number
95428
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.