Skip to content

Public FDA record

Z-0225-2015: The MectaLIF Oblique Handle, model number 03.22.10.0262, is used to insert the MectaLIF oblique spinal ca

Recall record Z-0225-2015 for Medacta Usa. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0225-2015
Company in FDA record
Medacta Usa
FDA classification date
Nov 18, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The MectaLIF Oblique Handle has the laser marking "MEDIAL" on the incorrect side and could result in incorrect insertion of the device during surgery.

Source facts

Action
Medacta sent an URGENT MEDICAL DEVICE RECALL letter dated October 21, 2014 were via e-mail to all sales representatives and distributors. The letters included instructions for the sales represetantives and distributors to: 1) check their inventory for the recalled devices; 2) immedaitely sequester the recalled devices and do not use them during surgery; 3) if the devices have been further distributed, inform them of the recall; and, 4) replacement devices will be provided. Customers with questions can contact Medacta USA at 805-910-6511.
Initiated
2014-10-21
Root cause
Employee error
Terminated
2016-03-14
Distribution
Worldwide Distribution - US including CA, CO, ID and Internationally to Australia and Germany.
Recall number
Z-0225-2015
Recall status
Terminated
CDRH record ID
131393
FDA classified
2014-11-18
Recall event number
69693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗