Skip to content

Public FDA record

Z-0225-2023: ORISE Gel Syringe Submucosal Lifting Agent Kits: indicated for use in gastrointestinal endoscopic procedu

Recall record Z-0225-2023 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0225-2023
Company in FDA record
Boston Scientific Corporation
FDA classification date
Nov 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention

Source facts

Action
***UPDATE 12/13/22** Boston Scientific issued Urgent Medical Device Product Removal Immediate Action Required Orise" Gel Submucosal Lifting Agent via Federal Express Priority Overnight mail on December 14, 2022, (as a Follow-Up To The Previous Product Advisory Dated October 10, 2022). Letter states reason for recall (Product Removal), health risk and action to take: Use and distribution of any remaining unused ORISE Gel Submucosal Lifting Agent products affected by this removal should cease immediately. For patients already treated with ORISE Gel, should the user identify submucosal distortions or mass formations in follow-up endoscopy, endoscopic ultrasound, imaging, or surgery, Boston Scientific recommends taking into account prior ORISE Gel use. Review pathology reports from the prior procedure to help determine the most appropriate course of action. Depending on the pathology present during the initial use of ORISE Gel, and whether it included conditions such as adenoma, high-grade dysplasia, or malignancy, a user may need to do nothing, repeat surveillance, repeat a biopsy, perform further mucosal resection, or plan surgical intervention to rule out any residual lesion. For each patient treated with ORISE Gel, Boston Scientific recommends appending their medical record with a copy of this letter to maintain awareness of this topic. If you identify any product from the affected lots within your inventory, please segregate the product immediately and return it to Boston Scientific in accordance with the enclosed Removal Instructions. If you are a distributor, please note that the depth of this removal is to the end user level and this notification should be forwarded to your customers. If you are a facility that has sent products to another hospital or facility within your network, please ensure that this notification is forwarded to them. Your local Sales Representative can answer any questions that you may have regarding this notification. _________________
Initiated
2022-10-10
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (Nationwide) and (Foreign) countries of: Algeria Argentina Argentina Australia Austria Bahrain Belgium Brazil Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Iraq Ireland Israel Italy Jordan Kuwait Latvia Lebanon Lithuania Luxembourg Mexico Netherlands New Zealand Norway Oman Palestine Panama Per¿ Poland Portugal Romania Russian Fed. Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Unit.Arab Emir. Zimbabwe
Recall number
Z-0225-2023
Recall status
Open, Classified
CDRH record ID
196429
FDA classified
2022-11-11
Recall event number
90986

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗