Skip to content

Public FDA record

Z-0225-2025: EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endos

Recall record Z-0225-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0225-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch

Source facts

Action
Olympus issued Urgent Medical Device Removal letter issued 9/13/24 to Endoscopy Department, Risk Management Department. Letter states reason for recall, health risk and action to take: 2. Examine your inventory and identify the above-listed device with the affected batch number. 3. Please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus. When it is received, you will receive a credit for your affected device(s). 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a) Go to https://olympusamerica.com/recall b) Enter the recall number "0453 c) Complete the form as instructed. Olympus requests that you report any complaints, including any incorrect labeling, to the Olympus Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). If you require additional information, please do not hesitate to contact me by e-mail at Cynthia.Ow@Olympus.com or by phone at (647) 999-3203.
Initiated
2024-09-13
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.
Recall number
Z-0225-2025
Recall status
Open, Classified
CDRH record ID
210135
FDA classified
2024-10-25
Recall event number
95400

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗