Z-0226-2022: Sensis/ Sensis Vibe systems with software VD1X as follows: (1)Sensis Model 10764561 (2) Sensis Vibe Hemo

Recall record Z-0226-2022 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0226-2022
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Nov 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs

Source facts

Action
Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice to customers via AX067/21/S dated 10/5/21. Letter states reason for recall,health risk and action to take: The software in the affected systems will be updated to display a notification message. Once the system has reached a runtime of more than 7 days, a notification message will be displayed after closing a study to remind the operator to reboot the system as it is running longer than advised. In addition, the system operating manual will include a warning message that the Sensis system can have performance issues when running for more than 7 days nonstop without a restart. A system running nonstop for more than 7 days may display incorrect or delayed vital signs; the values may not be saved, the report generation may be incorrect, and the system could become unavailable before or during an examination. The warning message will also contain the statement that the Sensis system should be restarted at least once a week. Siemens will provide a software update for all affected systems which will be distributed via Update Instruction AX066/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Initiated
2021-10-05
Root cause
Software design
Distribution
Nationwide
Recall number
Z-0226-2022
Recall status
Open, Classified
CDRH record ID
189845
FDA record created
2021-11-16
Recall event number
88870

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗