Skip to content

Public FDA record

Z-0226-2025: Boston Scientific myLUX Patient Application for an Apple Mobile device, Model Number 2939, Version 2.0.1

Recall record Z-0226-2025 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0226-2025
Company in FDA record
Boston Scientific Corporation
FDA classification date
Oct 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Patients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.

Source facts

Action
A field action/communication with Health Care Providers (HCPs) for affected patients was initiated on September 6, 2024, to inform them of the issue and provide the necessary patient data and connection support. Clinicians and patients were contacted via telephone calls. The BSC RhythmCARE team first called clinicians to describe the monitoring problem. With permission of the clinician, RhythmCARE Support contacted the patient to request a patient-initiated interrogation of the insertable cardiac monitor. Next, they offered options to troubleshoot and resolve the monitoring problem. In most cases, the troubleshooting included providing a BSC provided mobile monitor, in a singular case, the patients chose to buy a new mobile phone. The goal of these troubleshooting calls was to re-establish patient monitoring as scheduled. Recommendations were provided on an individual basis
Initiated
2024-09-06
Root cause
Under Investigation by firm
Distribution
Nationwide distribution to AL, AZ, CA, CO, FL, GA, ID, KY, MA, MD, NC, NJ, NY, OH, Puerto Rico, SC, UT, VA, WA.
Recall number
Z-0226-2025
Recall status
Open, Classified
CDRH record ID
210294
FDA classified
2024-10-25
Recall event number
95441

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗