Skip to content

Public FDA record

Z-0227-2023: Teleflex Medical RUSCH Pocket Pac i.c. Closed System Intermittent Catheter Kit, 14 Fr (4.7mm), REF 100961

Recall record Z-0227-2023 for TELEFLEX LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0227-2023
Company in FDA record
TELEFLEX LLC
FDA classification date
Nov 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product may not be sterile

Source facts

Action
Teleflex sent an URGENT MEDICAL DEVICE notice to its consignees on 10/03/2022 by Fed Ex 2nd day. The notice explained the problem and the hazard and requested the following actions be taken: "Action list number 1 Medical facilities 1. We request that you immediately check your inventory for product within the scope of this recall. Users should cease use and distribution of affected product and immediately quarantine the affected product. 2. If you have affected product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and contact Teleflex Customer Service at the contact details provided below. Teleflex Customer Service will issue a return goods authorization (RGA) number to you. Write the (RGA) number into the respective field in the Acknowledgement Form (Appendix 1) and promptly return this form by email to Teleflex Customer Service at the contact details provided below. You may then return all product in scope. 3. If you do not have affected product, mark the applicable checkbox on the Acknowledgement Form (Appendix 1) and return the form to Teleflex Customer Service at the contact details provided below. 4. Teleflex (or your local dealer) will issue a credit note upon receipt of the returned affected product. Action list number 2 Distributors 1. Provide this recall notice to all customers who have received affected product. Each customer is then required to complete the Acknowledgement Form (Appendix 1) and return it to you. 2. We request that you immediately check your inventory for affected product. Cease use and distribution of affected product and immediately quarantine the affected product. You may then return all product in scope. 3. As a distributor, you are then required to confirm to Teleflex that you have completed the field activity outlined in actions 1 and 2 of this Action List Number 2. Upon completion of your actions, please forward the completed Acknowledgement Form (Appendix 1) to Teleflex Customer Service at the conta
Initiated
2022-09-20
Root cause
Nonconforming Material/Component
Distribution
TN, CA, NC, TX, FL, OH, MO, IL, OR
Recall number
Z-0227-2023
Recall status
Open, Classified
CDRH record ID
196306
FDA classified
2022-11-11
Recall event number
90962

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗