Skip to content

Public FDA record

Z-0227-2025: Elekta Disposable Biopsy Needle Kit for Leksell Stereotactic System, REF 911933, Diameter / 14 Gauge, con

Recall record Z-0227-2025 for Elekta Instrument AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0227-2025
Company in FDA record
Elekta Instrument AB
FDA classification date
Nov 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Elekta has become aware that Disposable Biopsy Needles can contain some microscopic debris on the inside of the biopsy needle. The material in the debris is stainless steel, same material as the biopsy needle. No debris has been found on the outer parts of the biopsy needles.

Source facts

Action
Elekta issued an URGENT IMPORTANT FIELD SAFETY NOTIFICATION to its consignees on 069/23/2023 via email. Then notice explained the problem, clinical impact, and requested the destruction of the affected product. Contact the local Elekta representative for alternative solutions. They also requested that the notice be posted in a place accessible to all users until this action is closed and to notify all appropriate personnel working with this product of the content of the notice.
Initiated
2024-09-25
Root cause
Under Investigation by firm
Distribution
US: AZ, TX, FL, CA, NY, GA, VA, NJ, MO, PA, WA,
Recall number
Z-0227-2025
Recall status
Open, Classified
CDRH record ID
210361
FDA classified
2024-11-06
Recall event number
95484

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗