Skip to content

Public FDA record

Z-0228-2007: VerifyNow P2Y12 Assay Device Kit, Part Number 85054

Recall record Z-0228-2007 for Accumetrics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0228-2007
Company in FDA record
Accumetrics Inc
FDA classification date
Dec 6, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The VerifyNow P2Y12 assay can report an erroneous result instead of an error message when a sample with low hematocrit is run.

Source facts

Action
The 333 consignees were notified by regular mail or by E-mail on 9/5/2006 -- 9/13/2006. There were four versions of the notification letter tailored to (1) domestic customers; (2) domestic distributors; (3) international customers; and (4) international distributors. Each notification was accompanied by a 'fax back' form which customers were asked to sign and return to Accumetrics as an acknowledgement they received the notification letter. The notification letter instructs consignees that test results with the VerifyNow P2Y12 Assay and all VerifyNow assays should be interpreted in conjunction with other clinical and laboratory data available to the clinician. Caution is advised when testing patients outside the ranges documented in the package insert.
Initiated
2006-09-05
Root cause
Other
Terminated
2012-03-26
Distribution
Worldwide Distribution- USA, and countries of Australia, Canada, France, Germany, Italy, Japan, Korea, Malaysia, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, & United Kingdom
Recall number
Z-0228-2007
Recall status
Terminated
CDRH record ID
49194
FDA classified
2006-12-06
Recall event number
36707

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.