Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip toward the tip of the catheter and block the inflation hole, leading to the risk of the tip breaking off, and potentially becoming lost in the blood vessel.
Source facts
Action
On September 30, 2022, the recalling firm notified affected customers via Urgent Field Safety Notice letters.
Users were instructed to immediately identify, quarantine, and return the affected devices.
Initiated
2022-09-30
Root cause
Process design
Distribution
No US distribution. Foreign distribution to United Kingdom, Germany, France, Sweden, Austria, and Italy.
Recall number
Z-0228-2023
Recall status
Open, Classified
CDRH record ID
196484
FDA classified
2022-11-11
Recall event number
91006