Open original FDA record ↗Recall record
- FDA identifier
- Z-0229-04
- Company in FDA record
- Alcon Laboratories, Inc
- FDA classification date
- Dec 17, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm initiated the recall on 11/18/2003 via letters to all consignees. The letter advises physicians to stop using any patient bracelets (blue color) that contain the word 'stickoxid', destroy all blue bracelets, and identify all patients who had either ISPAN gas administered who may still have a bubble present in the eye.
Source facts
- Action
- The patient bracelet that is given to patients post surgery after administering either ISPAN Perfluoropropane or ISPAN Sulfur Hexafluordie referred to the term, 'stickoxid', which is the general term for nitrogen oxide and may be interpreted as either nitrogen monoxide (NO) or dinitrogen monxoide. The firm is recalling the old (blue) bracelets and replacing them with new (grey) bracelets that now reference the specific term dinitrogen monoxide N20 (laughing gas).
- Initiated
- 2003-11-18
- Root cause
- Other
- Terminated
- 2004-10-20
- Distribution
- Germany, Switzerland, and Austria
- Recall number
- Z-0229-04
- Recall status
- Terminated
- CDRH record ID
- 30459
- FDA classified
- 2003-12-17
- Recall event number
- 27748