Skip to content

Public FDA record

Z-0229-2025: Olympus Endoscopic Flushing Pump, Model/Catalog Number: OFP-2 (K10001141)

Recall record Z-0229-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0229-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 9/20/24 was sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. The range of affected serial numbers is included at the top of this letter. The serial number is on a label affixed to the back of the OFP-2 pump. An Olympus representative will create a Return Material Authorization for the return of your device to Olympus for an inspection. You may continue to use your OFP-2 pump until an Olympus representative performs the inspection. If the OFP-2 fails the inspection, the OFP-2 will require repair. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Olympus requests that you acknowledge receipt of this letter. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall b. Enter the recall number "0455 c. Complete the form as instructed. 3. If you have further distributed this product, identify your customers, and forward them this notification. 4. Beginning November 2024, please contact Olympus Customer Service at 1-800-848-9024, option 3, to request an Air Switch Assembly Inspection . Olympus will arrange for the return of your device to our Olympus Repair Center at a mutually convenient time. The Olympus representative s record of repair will also serve as the completion record of this action for your facility. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@Olympus.com.
Initiated
2024-08-23
Root cause
Equipment maintenance
Distribution
US Nationwide distribution.
Recall number
Z-0229-2025
Recall status
Open, Classified
CDRH record ID
210293
FDA classified
2024-10-25
Recall event number
95440

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗