Skip to content

Public FDA record

Z-0230-04: IMx Tacrolimus II Assay, list 3C10-20; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott Park

Recall record Z-0230-04 for Abbott Laboratories HPD/ADD/GPRD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0230-04
Company in FDA record
Abbott Laboratories HPD/ADD/GPRD
FDA classification date
Dec 17, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inadequate instructions for use resulting in calibration errors and the protein interference data in the package insert is not representative of assay performances.

Source facts

Action
Device correction recall initiated by letters dated 11/12/03. Letters were sent to all current IMx Tacrolimus II customers worldwide, advising them that calibration errors and controls out of range be the result of variance with the sample pretreatment steps listed in the package insert, and that the protein interference data in the insert is not representative of assay performance (12% when it should state less than 22% error) . A Customer Training Guide was included in the letter, outlining steps to aid the assay user in performing the pretreatment step and listing the most common deviations that can impact results.
Initiated
2003-11-13
Root cause
Other
Terminated
2004-10-29
Distribution
Nationwide and internationally to Guatemala, Venezuela, Columbia, Peru, Chile, Australia, Panama, Argentina, Puerto Rico, Dominican Republic, Canada, Germany, New Zealand, England, Japan, Singapore, Thailand, Korea and Costa Rica.
Recall number
Z-0230-04
Recall status
Terminated
CDRH record ID
30404
FDA classified
2003-12-17
Recall event number
27717

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗