Z-0230-2026: Fuse ULTRA Foot Plating System Instructions for Use. Models: OS200400, OS200401, OS328017, OS200020L, OS

Recall record Z-0230-2026 for CPM Medical Consultants, LLC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0230-2026
Company in FDA record
CPM Medical Consultants, LLC.
FDA classification date
Oct 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Instructions for use for a foot plating system is being updated to remove the second option of requirements for sterilization which was 3 minutes exposure time at 134-137-degrees C (273-279-degrees F) and 40mBar for 5-10 minutes drying time, because these requirements do not apply in the United States.

Source facts

Action
On 9/17/2025, correction notices were mailed to customers and distributors who were informed of the following: - The only sterilization requirement now shown on the new Rev D IFU is 4 mins exposure time at 132-Degrees C (270-Degrees F) and 20-30mins drying time. - All affected versions of the IFU should be immediately discarded. - You can access the e-IFU at: https://fusemedical.com/wp-content/uploads/2025/09/Fuse-ULTRA-Foot-Plating-System-IFU-Rev-D-9.2025.pdf - Please pass on this Medical Device Correction to all users of the products and all other people who need to be aware within your organization. - If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form, and updated IFU. - Complete and return the response form via email to FMohammad@FuseMedical.com If you have any questions, contact the firm at info@fusemedical.com
Initiated
2025-09-17
Root cause
Labeling design
Distribution
US Nationwide distribution in the states of NV, TX, LA, WA, WI, FL.
Recall number
Z-0230-2026
Recall status
Open, Classified
CDRH record ID
216331
FDA classified
2025-10-21
Recall event number
97679

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗