Open original FDA record ↗Recall record
- FDA identifier
- Z-0231-2018
- Company in FDA record
- Radiometer America Inc
- FDA record creation date
- Dec 8, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to misinterpretation of the barcode by the scanner, when the registration receipt barcode is scanned by the analyzer , a result from a different patient will be printed or displayed on the analyzer.
Source facts
- Action
- Customer were notified via letter using FedEx tracking on approximately October 24, 2017. Instructions included verification that the test results and the patient and demographic information matches, that Radiometer America is working to improve the software to remove the possibility of the code being misinterpreted, and to notify customers if the product has been further distributed. For further questions, please call (800) 736-0600.
- Initiated
- 2017-10-24
- Root cause
- Software in the Use Environment
- Terminated
- 2023-10-10
- Distribution
- USA (nationwide) to: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. Distributed internationally to Canada.
- Recall number
- Z-0231-2018
- Recall status
- Terminated
- CDRH record ID
- 159600
- FDA record created
- 2017-12-08
- Recall event number
- 78445