Z-0231-2023: Hill-Rom Progressa Bed System

Recall record Z-0231-2023 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0231-2023
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Nov 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The compression links on impacted Progressa beds, listed in Appendix A, may bend when raising the head section. The labeling for these beds list a maximum patient weight of 500 lbs., however the bending may still occur with a patient weight of less than 500 lbs. depending on the patient 's position on the bed.

Source facts

Action
Baxter/Hill-Rom disseminated an URGENT MEDICAL DEVICE CORRECTION NOTICE to its consignees on 10/18/2022 by UPS Ground Service. The notice explained the issue and potential risk and requested the following: "Actions to be taken by Customer: " Please identify if you have the impacted beds per Appendix A in your facility and remove from service immediately. Serial number can be found on the patient right side of the bed under the head rail. Return the enclosed response form within two weeks, confirming serial numbers of impacted beds in your facility. Action to be taken by the Distributor: " Please share this Medical Device Correction Notice with your end users. " Complete response form and return to hillrom2022-045@Sedgwick.com within two weeks, identifying serial numbers of beds to be reworked. Hillrom will arrange shipment of replacement kit and associated instructions." "If you have any questions regarding this Urgent Medical Device Correction, please contact Hillrom Technical Support at 800-445-3720 or technical.support@hillrom.com" Update: On 11/22/2022 (and some on 12/12/2022), the firm issued an URGENT MEDICAL DEVICE CORRECTION NOTICE to additional consignees via UPS Ground Service which provided the same instructions as the initial notification but added 59 additional serial numbers in the Appendix of the notification.
Initiated
2022-10-18
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, HI, IN, LA, MA, MO, NC, NM, NV, NY, OK, OR, PA, SC. TN, TX, WA, and WI. (UPDATED: States of PA and SC and there was government distribution.) There was no military distribution. Foreign distribution was made to Canada, Qatar, Singapore, and UAE. UPDATE: Saudi Arabia now included.
Recall number
Z-0231-2023
Recall status
Open, Classified
CDRH record ID
196422
FDA classified
2022-11-14
Recall event number
90990

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗