Skip to content

Public FDA record

Z-0231-2025: ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic

Recall record Z-0231-2025 for ImaCor Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0231-2025
Company in FDA record
ImaCor Inc.
FDA classification date
Oct 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

Source facts

Action
ImaCor issued Inventory Correction for ImaCor ClariTEE Probe (CLT-010 and CLT-010-1) letter on 9/16/24. Letter states reason for recall, health risk and action to take: 1. Identify the Affected Devices o Please check your inventory for ClariTEE probes with lot numbers 20250509 and 20250823. Inform ImaCor of the number of applicable probes currently in inventory. 2. Prepare the Products for Return o Package the affected products securely to prevent damage during shipment. 3. Shipping Instructions o Please ship the affected products via FedEx using FedEx Account Number 665380920 to cover the shipping cost. Ship to: ImaCor Inc, 50 Jericho Turnpike, Suite 105, Jericho, NY 11753. Destination phone number is (516) 393-0970. o Attach a copy of this letter inside the package for reference. 4. Replacement Devices o Once we receive the returned devices, we will ship replacement probes to your facility. Replacement probes are expected to be available on or about October 15, 2024. Please return all affected devices by October 7. If you have any questions or need assistance, please contact Jenn Kujawski at jenn@imacorinc.com. Our team is ready to assist you with any inquiries and can provide additional details if needed.
Initiated
2024-09-16
Root cause
Process control
Distribution
FL ,GA, NJ, TN
Recall number
Z-0231-2025
Recall status
Open, Classified
CDRH record ID
210334
FDA classified
2024-10-25
Recall event number
95459

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗