Open original FDA record ↗Recall record
- FDA identifier
- Z-0231-2026
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- Oct 22, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
Source facts
- Action
- An "URGENT MEDICAL DEVICE CORRECTION" notification letter dated 8/22/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your device in accordance with the instructions in the MAC VU360 Operator Manual and the actions described below: 1. Inspect all potentially affected devices for the presence of damage to the Acquisition Trunk Cable per the MAC VU360 Operator Manual (P/N 2088531-202, Section 11). 2. If the Acquisition Trunk Cable shows damage, remove the cable from service. 3. Ensure the Acquisition Module is placed correctly in the Acquisition Module Holder and that the Acquisition Trunk Cable is not pinched during insertion. See section 1.3 of the MAC VU360 V1.03 Setup and Configuration Manual (P/N 2088531-204). Note: Manuals can be accessed from the GE HealthCare Customer Documentation Portal at https://www.gehealthcare.com/support/manuals 4.Please complete and return the attached acknowledgement form to DCAR.Recall@gehealthcare.com GE HealthCare will provide a customer-installable kit containing a new Acquisition Trunk Cable and a redesigned Acquisition Module Holder for the affected devices at no cost to you. Please make sure all potential users in your facility are made aware of this safety notification and the recommended actions. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
- Initiated
- 2025-08-22
- Root cause
- Device Design
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
- Recall number
- Z-0231-2026
- Recall status
- Open, Classified
- CDRH record ID
- 215930
- FDA classified
- 2025-10-22
- Recall event number
- 97564