Open original FDA record ↗Recall record
- FDA identifier
- Z-0232-2016
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Nov 6, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte
Source facts
- Action
- One consignee was contacted via a Boston Scientific "Medical Device Retrieval" letter dated September 17, 2015. The letter was addressed to Hospital Administrator. The letter described the problem and the product involved in the recall. Informed consignee that their local Boston Scientific sales representative will be retrieving the product form their inventory. For questions they can contact their local sales representative or Boston Scientific Technical service at 1-800-227-3422. A second Boston Scientific "Important Medical Device Information" letter dated October 2015 was hand delivered to physicians starting on 10/15/2015. The letter described the problem and provided information on Clinical Considerations, Affected Population, Recommendations and Further Information. For questions they can contact their Boston scientific representative or the Americas Technical service at 1-800-227-3422.
- Initiated
- 2015-09-17
- Root cause
- Process design
- Terminated
- 2016-03-08
- Distribution
- Distributed in the US to MA and PR.
- Recall number
- Z-0232-2016
- Recall status
- Terminated
- CDRH record ID
- 140276
- FDA classified
- 2015-11-06
- Recall event number
- 72225