Open original FDA record ↗Recall record
- FDA identifier
- Z-0232-2018
- Company in FDA record
- Fujifilm Medical Systems U.S.A., Inc.
- FDA record creation date
- Dec 10, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.
Source facts
- Action
- FUJIFILM Medical Systems U.S. A. Inc. (FMSU) sent "Urgent: Medial Device Correction" recall notification letter dated June 21, 2017 to all Customers. Customers were notified of the recall, potential issue, potential hazard, description of the issue, the actions planned to correct the issue, the actions to be taken by the customer and contact information. A Customer Feedback Form was provided accompanying the letter. FMSU service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures. As FMSUs planned field correction at the affected sites, FMSU will install the additional protective parts preventing the X-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. For further questions, please call (203) 324-2000.
- Initiated
- 2017-06-15
- Root cause
- Device Design
- Terminated
- 2018-05-09
- Distribution
- US Distribution.
- Recall number
- Z-0232-2018
- Recall status
- Terminated
- CDRH record ID
- 160059
- FDA record created
- 2017-12-10
- Recall event number
- 78644