Skip to content

Public FDA record

Z-0232-2018: FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name)

Recall record Z-0232-2018 for Fujifilm Medical Systems U.S.A., Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0232-2018
Company in FDA record
Fujifilm Medical Systems U.S.A., Inc.
FDA record creation date
Dec 10, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Source facts

Action
FUJIFILM Medical Systems U.S. A. Inc. (FMSU) sent "Urgent: Medial Device Correction" recall notification letter dated June 21, 2017 to all Customers. Customers were notified of the recall, potential issue, potential hazard, description of the issue, the actions planned to correct the issue, the actions to be taken by the customer and contact information. A Customer Feedback Form was provided accompanying the letter. FMSU service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures. As FMSUs planned field correction at the affected sites, FMSU will install the additional protective parts preventing the X-ray tube assembly from coming into contact with the patient or operator should separation of the tube mounting flange from the tube holding shaft occur. For further questions, please call (203) 324-2000.
Initiated
2017-06-15
Root cause
Device Design
Terminated
2018-05-09
Distribution
US Distribution.
Recall number
Z-0232-2018
Recall status
Terminated
CDRH record ID
160059
FDA record created
2017-12-10
Recall event number
78644

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗