Skip to content

Public FDA record

Z-0232-2023: BD Phoenix PMIC/ID-107: The Phoenix PMIC/ID panel is used for the rapid identification and susceptibilit

Recall record Z-0232-2023 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0232-2023
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Nov 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Misformulated minocycline-can cause erroneously increased Minimal Inhibitory Concentration (MIC) values when performing antibiotic susceptibility testing (AST) against minocycline and result in the reporting of false resistance for clinical samples if the initial Quality Control (QC) testing of the panel batch passes, delaying treatment of the patient with the most appropriate antibiotic

Source facts

Action
BD initiated the recall on October 13, 2022 via Urgent Medical Device Product Advisory states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above and ensure the contents of this Product Advisory are read and understood. 2. It is recommended to suspend reporting of Minocyline AST results to clinicians for the affected lot. 3. Share and post this Product Advisory letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. BD Contact US Contact Information Areas of Support North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Initiated
2022-10-13
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0232-2023
Recall status
Open, Classified
CDRH record ID
196503
FDA classified
2022-11-14
Recall event number
91010

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.