Skip to content

Public FDA record

Z-0232-2026: t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014

Recall record Z-0232-2026 for Tandem Diabetes Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0232-2026
Company in FDA record
Tandem Diabetes Care, Inc.
FDA classification date
Oct 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

Source facts

Action
On 8/5/2025, Field Safety Notices were sent to customers informing them of the following: Actions to avoid the issue: - Wait until your G7 sensor session ends before inserting a new G7 sensor. - However, if you do insert a new G7 sensor before your last sensor session ends, wait to pair until at least 20 minutes after the last session ends. Firm is working on a software update to resolve this issue. If you have questions or need help, please call firm's Customer Support 1-877-801-6901 or email Techsupport@tandemdiabetes.com. Our team is available 24/7/365. On 9/16/2025, correction notices were emailed to customers who were informed the firm updated user guides to align with instructions in the correction notice. Also, the firm will notify impacted users via an updated Field Correction Notice when the software update is available. The notice will provide instructions on how to remotely update their software.
Initiated
2025-08-05
Root cause
Software change control
Distribution
US nationwide distribution including Puerto Rico.
Recall number
Z-0232-2026
Recall status
Open, Classified
CDRH record ID
215511
FDA classified
2025-10-23
Recall event number
97403

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗