Open original FDA record ↗Recall record
- FDA identifier
- Z-0233-2024
- Company in FDA record
- Unomedical A/S
- FDA classification date
- Nov 21, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.hyperglycemia). Diabetic ketoacidosis.
Source facts
- Action
- Unomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
- Initiated
- 2023-10-11
- Root cause
- Process change control
- Distribution
- US Nationwide distribution in the state of CA.
- Recall number
- Z-0233-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203810
- FDA classified
- 2023-11-21
- Recall event number
- 93269