Open original FDA record ↗Recall record
- FDA identifier
- Z-0233-2026
- Company in FDA record
- Baxter Healthcare Corporation
- FDA classification date
- Oct 23, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Source facts
- Action
- Beginning September 17, 2025, firm notified impacted distributors via Medical Device Recall letters. Customers were instructed to return any affected product and submit a request for replacements. Distributors were instructed to notify their end users via a provided Important Product Information Letter. For general questions regarding this action, please contact Welch Allyn technical service at 800-535-6663. Select the following sequence of options: 2 (technical support), 1(English), 1 (physical exam devices), 3 (thermometry), between the hours of 8:00 AM and 8:00 PM Eastern Time, Monday through Friday.
- Initiated
- 2025-09-17
- Root cause
- Process control
- Distribution
- Distribution US nationwide, Canada and Colombia.
- Recall number
- Z-0233-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216117
- FDA classified
- 2025-10-23
- Recall event number
- 97614