Skip to content

Public FDA record

Z-0234-2015: Etest Ceftaroline (CPT32) US B30. Etest (Antimicrobial Susceptibility Testing) is a quantitative techniq

Recall record Z-0234-2015 for Biomerieux. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0234-2015
Company in FDA record
Biomerieux
FDA classification date
Nov 19, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Source facts

Action
Consignees of bioMrieux, impacted by the identified issue, were notified of the FSCA and corrective action (FSCA notification) electronically on October 16th, 2014. Additionally, the bioMrieux United States subsidiary notified US consignees of the FSCA notification via FedEx on October 23rd, 2014.
Initiated
2014-10-16
Root cause
Labeling Change Control
Terminated
2017-11-30
Distribution
Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.
Recall number
Z-0234-2015
Recall status
Terminated
CDRH record ID
130987
FDA classified
2014-11-19
Recall event number
69579

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗