Open original FDA record ↗Recall record
- FDA identifier
- Z-0235-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Nov 20, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Siemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that simulation system components could collide with patient and/or system equipment.
Source facts
- Action
- Customers were initially notified about the issue by letter on January 7, 2014 (this action was reported as Z-0786-2014 which has been terminated). Siemens promised to notify them when a fix became available. Customer letter, dated November 7, 2014, was sent to notify end users that a fix has become available and Siemens would contact them to schedule an appointment for the correction.
- Initiated
- 2014-10-10
- Root cause
- Device Design
- Terminated
- 2017-06-29
- Distribution
- Worldwide Distribution - US Nationwide in the states of PA, KY, LA, MA, MI, and CA; and countries of: Argentina, Belarus, Bulgaria, Costa Rico, P.R. China, Curacao, Egypt, Repulic Korea, Germany, Greece, Hungary, Italy, Iraq, Jordan, Malaysia, Palestanian Ter, Poland, Thailand, Turkey, and Vietnam.
- Recall number
- Z-0235-2015
- Recall status
- Terminated
- CDRH record ID
- 131032
- FDA classified
- 2014-11-20
- Recall event number
- 69601