Beckman Coulter has identified that due to a software nonconformity in connection with a changed service setting Remisol could display wrong results.
Source facts
Action
Beckman Coulter sent an Urgent Medical Device Recall letter dated October 16, 2017, to all affected customers.
Beckman Coulter has identified a potential issue due to a software nonconformity which could occur after a default service setting (NoCondition) is manually changed from off to on, potentially causing results outside of analytical range to be erroneously displayed.
For further questions, please call (800) 526-3821.
Initiated
2017-10-16
Root cause
Software design
Terminated
2024-02-13
Distribution
Worldwide Distribution - US Distribution and to the countries of : Australia, France, Germany, Israel and Russia
Recall number
Z-0235-2018
Recall status
Terminated
CDRH record ID
159406
FDA record created
2017-12-13
Recall event number
78394