Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical
interference between the device fan and filter
Source facts
Action
Stryker notified accounts by letter on 9/6/18 via Fed'X advising of the problem, steps to be taken: 1.Locate the units listed on the attached business reply form and identify the address where they can be serviced. 2.Return the enclosed business reply form to confirm receipt of this notification by fax 269 488 8691 or email productfieldaction@stryker.com. 3.Upon receipt of the completed business reply form, Stryker will contact you to arrange for the modification/repair of your Altrix unit. 4.If you have questions regarding the safe continued use of your Altrix unit, please contact Tech Support by calling +1 800 327 0770 and selecting opt 2.
Initiated
2018-09-06
Root cause
Software design
Terminated
2020-04-13
Distribution
Worldwide - US Nationwide Distribution and the countries of Foreign: Australia, Canada, Netherlands, Switzerland and United Kingdom
Recall number
Z-0235-2019
Recall status
Terminated
CDRH record ID
167452
FDA record created
2018-10-24
Recall event number
81103