Z-0235-2019: Altrix Precision Temperature Management System Model: 8001Product Usage: The Altrix system is intended fo

Recall record Z-0235-2019 for Stryker Medical Division of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0235-2019
Company in FDA record
Stryker Medical Division of Stryker Corporation
FDA record creation date
Oct 24, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical interference between the device fan and filter

Source facts

Action
Stryker notified accounts by letter on 9/6/18 via Fed'X advising of the problem, steps to be taken: 1.Locate the units listed on the attached business reply form and identify the address where they can be serviced. 2.Return the enclosed business reply form to confirm receipt of this notification by fax 269 488 8691 or email productfieldaction@stryker.com. 3.Upon receipt of the completed business reply form, Stryker will contact you to arrange for the modification/repair of your Altrix unit. 4.If you have questions regarding the safe continued use of your Altrix unit, please contact Tech Support by calling +1 800 327 0770 and selecting opt 2.
Initiated
2018-09-06
Root cause
Software design
Terminated
2020-04-13
Distribution
Worldwide - US Nationwide Distribution and the countries of Foreign: Australia, Canada, Netherlands, Switzerland and United Kingdom
Recall number
Z-0235-2019
Recall status
Terminated
CDRH record ID
167452
FDA record created
2018-10-24
Recall event number
81103

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗