It was discovered that a software defect may result in the scanner not terminating the CT scan at the intended location.
Source facts
Action
Philips Healthcare is directly notifying affected users via letter. In addition, Philips Healthcare has performed the software correction for forward production and plans a Field Correction to fix the defect in the field. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office.
Initiated
2014-07-31
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-02-16
Distribution
US Distribution
Recall number
Z-0236-2015
Recall status
Terminated
CDRH record ID
130286
FDA classified
2014-12-16
Recall event number
69356