Open original FDA record ↗Recall record
- FDA identifier
- Z-0236-2019
- Company in FDA record
- GE Medical Systems, LLC
- FDA record creation date
- Oct 24, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.
Source facts
- Action
- On about 10/03/2018, GE Healthcare sent URGENT MEDICAL DEVICE CORRECTION letters to their consignees informing them of recall and providing safety instructions. Consignees were instructed that they can continue to use the system by following the existing Operator Manual Safety Information for proper patient padding and observation during scanning. The safety information contained in the Operator's Manual includes: 1. WARNING: Place appropriate non-conductive padding between the patient and the bore wherever a portion of the body may come into contact with the magnet opening. 2. CAUTION: Continuous patient observation and contact are required in all modes of operation. 3. CAUTION: Extra attention should be utilized when scanning patients who are unconscious, sedated, or may have loss of feeling in any body part or are physically or mentally impaired and unable to alert the personnel. If a brown discoloration is seen on the inner surface of the patient bore, or if a burning smell is noticed, please contact their local Service Representative. GE Healthcare will contact consignees to arrange for the correction to the devices. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
- Initiated
- 2018-10-03
- Root cause
- Device Design
- Terminated
- 2022-05-27
- Distribution
- Worldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.
- Recall number
- Z-0236-2019
- Recall status
- Terminated
- CDRH record ID
- 168339
- FDA record created
- 2018-10-24
- Recall event number
- 81348