Z-0236-2022: BIO-RAD BioPlex 2200 REF 12000651 Syphilis Total & RPR Calibrator Set CAL The BioPlex 2200 Syphilis Tot

Recall record Z-0236-2022 for Bio-Rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0236-2022
Company in FDA record
Bio-Rad Laboratories, Inc.
FDA record creation date
Nov 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to two issues: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure.

Source facts

Action
On 09/17/2021, the Recalling Firm sent an "URGENT PRODUCT NOTIFICATION" Letter to customer via overnight FedEx informing them of two issues their Syphilis Total and RPR (Rapid Plasma Reagin) kits: 1) Customer complaints associated with greater than expected Rapid Plasma Reagin (RPR) reactivity which does not confirm with treponemal antibody test which may be due to a positive correlation between the mass COVID vaccinations and false reactive RPR results, and 2) During routine stability monitoring, an increase in mean bias above specification was observed which can cause elevated RPR reactivity potentially due to early shelf-life failure. Customer are instructed to: -Immediately discontinue use and dispose of any unused product of reagent packs and calibrator set according to local waste management procedures. -Complete the customer response form and return it to their local Recalling Firm Technical Support, or return this form to Techsupport.ussd@bio-rad.com or the Toll-Free Fax 1-888-228-0688 within 10 days of receiving this notice.. -If the affected products have been further distributed or transferred, notify those customers of the field action. For additional questions, contact local Bio-Rad Technical Support.
Initiated
2021-09-17
Root cause
Nonconforming Material/Component
Distribution
Worldwide Distribution: U.S.(nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, and WV; O.U.S.(foreign) countries of: France, Germany, Israel, Italy, Russian Federation, South Africa, Spain, and Taiwan.
Recall number
Z-0236-2022
Recall status
Open, Classified
CDRH record ID
189880
FDA record created
2021-11-16
Recall event number
88874

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗