Reason stated in the source
The device is missing grit blasting on the proximal surface of the femoral stem, which may lead to the possibility of product revision in the event of undersizing.
Source facts
- Action
- Affected agents and representatives were notified by email, with initial communication sent via a quarantine form on 10/1/21. Additional notices titled "Urgent Medical Device Recall" were emailed out on October 15, 2021. Customers were instructed to immediately quarantine lot 38933 and return the product to OMNI. Customers were provided with a response form and asked to complete and return the form to the recalling company by FAX to 508-819-3390 or my email at christina.rovaldi@coringroup.com. For further information contact Christina Rovaldi, RA/QA Manager at 774-226-1847 or 800-448-6664. Company contact hours are Monday through Friday, 8:00 AM to 4:30 PM, Eastern Time.
- Initiated
- 2021-10-01
- Root cause
- Process control
- Terminated
- 2022-07-20
- Distribution
- US Distribution to states of: Colorado and Virginia.
- Recall number
- Z-0237-2022
- Recall status
- Terminated
- CDRH record ID
- 189951
- FDA record created
- 2021-11-16
- Recall event number
- 88890