Open original FDA record ↗Recall record
- FDA identifier
- Z-0237-2023
- Company in FDA record
- Whele LLC DBA Perch
- FDA classification date
- Nov 23, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The products may have the potential to burn customers. Firm has received complaints of burns, mild shocks, and rashes/irritation.
Source facts
- Action
- On October 24, 2022, the firm notified retailers Amazon.com and Walmart of the recall. These internet retailers were instructed to notify users of the product about the recall using their customer information. Whele LLC will also have a website for the public that will include instructions for the recall as well as customer support for the recall. The website can be found at https://www.mightyblissheatingpadrecall.expertinquiry.com/ Users are asked to immediately cease use of the product and safely destroy the product by cutting the cord (after ensuring the product is unplugged.) Users are asked to take a photo of the destroyed product and upload it to the firm's recall website to verify that the product can no longer be used. Firm also issued press release: https://www.prnewswire.com/news-releases/whele-llc-announces-national-voluntary-recall-of-mighty-bliss-electric-heating-pad-due-to-product-safety-concerns-301656419.html
- Initiated
- 2022-10-24
- Root cause
- Device Design
- Distribution
- US Nationwide through Amazon.com and Walmart.
- Recall number
- Z-0237-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196373
- FDA classified
- 2022-11-23
- Recall event number
- 90977