Open original FDA record ↗Recall record
- FDA identifier
- Z-0237-2024
- Company in FDA record
- CHANGE HEALTHCARE CANADA COMPANY
- FDA classification date
- Nov 6, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Change Healthcare has identified a software issue where Change Healthcare Radiology Solutions 14.0, 14.1, and 14.2 may not update records in external applications.
Source facts
- Action
- Field Safety Notice was distributed to all consignees on10/12/2023 by email. The notice explained the issue, potential risk to the patient, and directed users to to check the CHRS Exception Tracker on a regular basis, as failed updates will be listed with details on the failure. Recommended actions to take on potentially affected data: Impacted data can be identified by using the CHRS Exception Tracker. Customers are advised to review records of recent updates from CHRS to external systems and manually update any changes to study records that may have been missed. Change Healthcare will install a software update.
- Initiated
- 2023-10-12
- Root cause
- Software design
- Distribution
- US, Canada, Ireland, Saudi Arabia, France, UK
- Recall number
- Z-0237-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203781
- FDA classified
- 2023-11-06
- Recall event number
- 93259